Ubicación: Global + English
Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Plastic Haze Meter Testing: Principles, Standards, and Applications

Plastic haze meter testing is a key optical measurement used to quantify light scattering and luminous transmittance in transparent and translucent plastic materials such as films, sheets, and molded components. Performed in accordance with internationally recognized standards including ASTM D1003, ISO 14782, and GB/T 2410, this test is essential for quality control in the packaging, automotive, electronics, and display industries. This article explains the measurement principle, applicable standards, sample requirements, testing procedures, and typical applications of haze meter testing for plastics.

Technical Overview

Haze is defined as the percentage of transmitted light that, when passing through a specimen, deviates from the direction of the incident beam by more than an average angle of 2.5 degrees due to forward scattering. A haze meter (hazemeter) is a photoelectric instrument designed to measure this scattered light component together with total luminous transmittance, providing two of the most important optical quality indicators for transparent plastics.

The instrument consists of a stabilized light source, a collimating optical system, an integrating sphere, a movable light trap, and a photodetector whose spectral response matches the photopic sensitivity of the human eye under CIE Standard Illuminant A. During measurement, four optical signals are collected: the incident light (T1), the total light transmitted through the specimen (T2), the light scattered by the instrument itself (T3), and the light scattered by both the instrument and the specimen (T4). Haze is then calculated as the ratio of diffuse transmittance to total transmittance:

Haze (%) = (T4/T2 − T3/T1) × 100, while total luminous transmittance is calculated as Tt (%) = T2/T1 × 100. Modern hazemeters achieve repeatability better than ±0.1 haze units, making them suitable for both laboratory certification and production-line quality control. The principal standards governing this measurement include ASTM D1003 (Standard Test Method for Haze and Luminous Transmittance of Transparent Plastics), ISO 14782 (Plastics — Determination of Haze for Transparent Materials), GB/T 2410 (Determination of Light Transmittance and Haze of Transparent Plastics), and the related JIS K 7361-1 procedure.

Testing Samples

Plastic Films and Flexible Packaging

Thin flexible films of PE, PP, BOPP, PET, PVC, PA, and PLA with thicknesses typically ranging from 15 µm to 200 µm are the most frequently tested sample type. Surface quality is critical: specimens must be free of wrinkles, creases, fingerprints, dust, and scratches, since any surface defect artificially increases the measured haze value.

Rigid Sheets and Plates

PMMA (acrylic), PC (polycarbonate), PS, SAN, and PETG sheets with thicknesses from 0.5 mm up to 10 mm or more can be measured, provided the specimen fits the instrument aperture. Because haze is strongly thickness-dependent, the reported result must always state the specimen thickness.

Molded and Optical Components

Injection-molded lenses, light guide panels, diffuser plates, instrument covers, and other flat transparent parts can be tested directly on their functional surfaces. Curved or textured parts should be measured on flat reference coupons molded under identical conditions whenever geometry prevents flush contact with the aperture.

Other Transparent Materials

The same instrumentation is widely applied to glass, coatings, adhesives, and transparent laminates, although the tolerances and acceptance criteria are defined by the material-specific standards for each product family.

Testing Items

Standard haze meter testing reports the following optical parameters:

  • Haze (wide-angle scattering): the percentage of transmitted light scattered more than 2.5° from the incident beam, which quantifies the milky or cloudy appearance of the material.
  • Total luminous transmittance (Tt): the percentage of incident light transmitted through the specimen in all forward directions, reflecting overall brightness or see-through performance.
  • Diffuse transmittance (Td): the absolute percentage of transmitted light that is scattered, obtained as the product of haze and total transmittance (Td = Haze × Tt / 100).
  • Specular transmittance: the fraction of light passing without measurable deviation, which can be derived from total and diffuse components when required by the specification.

Testing Methods

Specimen Preparation and Conditioning

Specimens are cut to cover the instrument aperture completely, typically 50 mm × 50 mm or larger, with a minimum of three specimens per sample for averaging. Prior to measurement, specimens are conditioned at 23 °C ± 2 °C and 50% ± 10% relative humidity in accordance with ISO 291 or ASTM D618 for a minimum of 40 hours. Both surfaces are cleaned with a lint-free wipe and, where appropriate, a suitable cleaning solvent, since contamination is the most common cause of false haze readings.

Instrument Calibration

The hazemeter is zeroed and calibrated using certified reference standards supplied with traceable calibration values for haze and transmittance. Calibration is verified before each test session, and the light trap alignment is checked to ensure that the 2.5° scattering geometry defined by the standards is maintained.

Measurement Procedure (ASTM D1003 Method A)

Under Method A, the four signals T1, T2, T3, and T4 are acquired automatically: first with no specimen in place to record the incident and instrument-scatter baselines, then with the specimen mounted flush against the integrating sphere aperture to record total and combined scattering. The instrument computes haze and transmittance from the standardized equations. Each specimen is measured at several positions, and the mean, standard deviation, and specimen thickness are reported.

Alternative Procedures

ASTM D1003 Method B permits measurement using a spectrophotometer with an integrating sphere, calculating haze from spectral data. ISO 14782 and GB/T 2410 describe equivalent integrating-sphere procedures with slightly different geometric tolerances, and JIS K 7361-1 aligns closely with the ISO method. When comparing results between laboratories, the same standard and the same specimen thickness must be specified, since haze values obtained under different methods are not directly interchangeable.

Expression of Results

The test report should include the material identification, specimen thickness, number of specimens, conditioning environment, the standard and method used, the individual and mean haze and transmittance values, and the instrument model and calibration date. This documentation ensures traceability and comparability of results between accredited third-party laboratories and in-house quality departments.

Applications

Haze meter testing supports material development, incoming inspection, and product certification across a broad range of industries:

  • Flexible and rigid packaging: verifying the clarity of BOPP, PET, and PE films for retail display and printing applications.
  • Automotive: evaluating instrument cluster covers, headlamp lenses, glazing, and interior trim panels against optical acceptance limits.
  • Displays and electronics: controlling the haze of protective cover films, touch panel overlays, and optical adhesives used in displays.
  • Lighting: characterizing diffuser plates and light guide panels where a precisely defined balance of haze and transmittance is required.
  • Construction and agriculture: assessing greenhouse films, polycarbonate roofing, and window glazing for daylighting performance.
  • Medical devices: confirming the transparency of packaging, infusion components, and diagnostic consumables for visual inspection purposes.

FAQ

What is the difference between haze and transmittance?

Transmittance measures how much light passes through a specimen in total, while haze measures how much of that transmitted light is scattered away from the original beam direction by more than 2.5 degrees. A material can have high transmittance but high haze (bright yet cloudy) or low haze with low transmittance (clear but dark), so both parameters must be reported together.

Which standards are used for plastic haze testing?

The most widely applied standards are ASTM D1003, ISO 14782, GB/T 2410, and JIS K 7361-1. All are based on integrating-sphere optics with a 2.5° scattering threshold and CIE Standard Illuminant A detection.

What sample size and thickness are required?

Specimens must fully cover the instrument aperture, typically 50 mm × 50 mm minimum, and be flat, clean, and free of surface defects. There is no fixed thickness limit, but thickness must be reported because haze increases with optical path length.

Why did my haze result come out higher than expected?

Common causes include surface contamination, fingerprints, static-attracted dust, scratches, poor conditioning, and excessive specimen thickness. Cleaning both surfaces and re-measuring at several positions usually distinguishes genuine material haze from surface artifacts.

How often should a haze meter be calibrated?

Instruments should be verified against certified reference standards before each testing session, with full traceable calibration performed at the interval recommended by the manufacturer or required by the laboratory quality system, typically annually.

Can tinted or colored plastics be tested?

Yes, since the detector operates over the full photopic range under Illuminant A. However, strongly colored or light-absorbing materials yield low transmittance values, and the reporting standard should be confirmed with the client or specification owner before testing.

Why Choose ZKGX?

  • State-of-the-art analytical equipment
  • Highly qualified scientific team
  • Fast turnaround time
  • Competitive pricing